- 15 -
6) Select “DICOM CONFORMANCE TEST” from MENU 1/2 using the “▼” or “▲” button
and press the “OK” button. Then select “Yes” for “START TESTING” from “DICOM
CONFORMANCE TEST” OSD and press the “OK” button.
Note
If “DICOM GSDF” is NOT selected, the message “START TESTING” will not
appear.
DICOM CONF ORMA NCETEST
RESULT :---
MA X DE VIATION : ---
LUMI NANCE :Stable
EL APSED TIME :123 m in
STA RT TE S TI N G
Y e s
>N o
DICOM CONFORMANCE TEST
START TESTING
Completing the measurement takes between 1 minute to 2 minutes.
7) Showing the te results
When the measurement is completed, the te results are shown.
DICOM CONF ORMA NCETEST
RESULT :Pa ssed
MA X DE VIATION : 4. 5
LUMI NANCE :Stable
EL APSED TIME :123 m in
STA RT TE S TI N G
Y e s
>N o
%
5MP
TEST RESULTS
Test results
Passed: Max.
deviation ≤ 10 %
Failed: Max. deviation > 10 %
Cancelled: Test has been
cancelled.
Passed(!), Failed(!)
When “Passed” or “Failed” is
followed by “(!)”, the test has
been made under not
recommended condition.
(The test was performed in less
than 60 minutes of continuous
lighting time of the backlight.)
If the test is continuously
performed in less than 30
seconds, “Unstable” is shown in
LUMINANCE and “(!)” is shown in
RESULT.
Maximum deviation
The test results are saved until the next test or calibration.
If the achieved maximum deviation does not conform to the standard even though
the test has been performed under the recommended condition, calibrate the monitor
using an external calibration sensor.
Without the OSD operation for over 15 seconds, the OSD disappears automatically.