5-4 Disassembly Model 2701 Service Manual
Disassembly procedures
Case cover removal
Follow the steps below to remove the case cover to gain access to internal parts.
WARNING Before removing the case cover, disconnect the line cord and any test leads
from the instrument.
1. Remove Handle — The handle serves as an adjustable tilt-bail. Adjust its position by
gently pulling it away from the sides of the instrument case and swinging it up or down.
To remove the handle, swing the handle below the bottom surface of the case and back
until the orientation arrows on the handles line up with the orientation arrows on the
mounting ears. With the arrows lined up, pull the ends of the handle away from the
case.
2. Remove Mounting Ears — Remove the screw that secures each mounting ear. Pull
down and out on each mounting ear.
NOTE When re-installing the mounting ears, make sure to mount the right ear to the right
side of the chassis and the left ear to the left side of the chassis. Each ear is marked
“RIGHT” or “LEFT” on its inside surface.
3. Remove Rear Bezel — To remove the rear bezel, loosen the two captive screws that
secure the rear bezel to the chassis. Pull the bezel away from the case.
4. Removing Bottom Screws — Remove the four bottom screws that secure the case to the
chassis.
5. Remove Cover — To remove the case, grasp the front bezel of the instrument and care-
fully slide the chassis forward. Slide the chassis out of the metal case.
Motherboard removal
Perform the following steps to remove the motherboard. This procedure assumes that the
case cover is already removed.
1. Remove the RS-232 and Digital I/O fasteners.
The RS-232 and Digital I/O connectors each have two nuts that secure the connectors to
the rear panel. Remove these nuts.
2. Remove the front/rear inputs switch rod.
At the switch, place the edge of a flat-blade screwdriver in the notch on the pushrod.
Gently twist the screwdriver while pulling the rod from the shaft.
3. Disconnect the front input terminals.
You must disconnect these input terminal connections:
INPUT HI and LO
SENSE HI and LO
AMPS

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm