EasyManua.ls Logo

Keithley 2701

Keithley 2701
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2701 Service Manual Calibration Reference B-19
:DATE?
(:CALibration:PROTected:CARD1:DATE?)
(:CALibration:PROTected:CARD2:DATE?)
Purpose To request the Model 7700 calibration date.
Format :cal:prot:card1:date?
:cal:prot:card2:date?
Response <year>, <month>, <day>
Description The :CARD1:DATE? and :CARD2:DATE? queries allow you to read back
the calibration date from a Model 7700 in Slot 1 and Slot 2, respectively.
NOTE The card calibration date is automatically set to the Model 2701 real time clock date
when the card is calibrated.
Example :CAL:PROT:CARD1:DATE? Request card 1 cal date.
:INIT
(:CALibration:PROTected:CARD1:INITiate)
Purpose To initiate Model 7700 calibration procedures.
Format :cal:prot:card1:init
Parameter None
Description The :INIT command enables Model 7700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the :CODE command, but before performing Model 7700
calibration.
Example :CAL:PROT:CARD1:INIT Initiate 7700 calibration.

Table of Contents

Other manuals for Keithley 2701

Related product manuals