1-10 Performance Verification Model 2701 Service Manual
Verifying AC voltage
Check AC voltage accuracy by applying accurate AC voltages at specific frequencies from
the AC voltage calibrator to the Model 2701 inputs and verifying that the displayed readings
fall within specified ranges.
CAUTION
Do not exceed 1000V peak between front terminals INPUT HI and
INPUT LO, or 8
×
10
7
V•Hz input, because instrument damage may occur.
Follow these steps to verify AC voltage accuracy:
1. Connect the Model 2701 HI and LO INPUT jacks to the AC voltage calibrator as shown
in Figure 1-2. Be sure the INPUTS switch is in the FRONT position.
Figure 1-2
Connections for Model 2701 AC volts verification
!
Model 2701 Ethernet Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT ENTER
DIGITS RATE
RELFILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
OUTPUT
RATIO
ACV
ACI
2 4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS ON/OFFDELAY
HOLD
SAVE SETUP
CONFIG HALT
TYPE
LSYNC
TEST
MONITOR
STEP SCAN
OCOMP
CH-OFF CARD
CLOSE
Integra Series
RS-232ETHERNET
Calibrator (Output AC Voltage)
Shielded
cable
Amplifier (Connect to calibrator)
Note: Amplifier required only
for 700V, 50kHz output.
Model 2701
OUTPUT
HI
OUTPUT
LO
INPUT HI
INPUT LO
CAT I

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals