B-20 Calibration Reference Model 2701 Service Manual
:LOCK
(:CALibration:PROTected:CARD1:LOCK)
Purpose To lock out Model 7700 calibration.
Format :cal:prot:card1:lock
Parameter None
Description The :LOCK command allows you to lock out Model 7700 calibration after
completing the procedure. Thus, :LOCK performs the opposite of enabling
calibration with the :CODE command.
Example :CAL:PROT:CARD1:LOCK Lock out card 1 calibration.
:LOCK?
(:CALibration:PROTected:CARD1:LOCK?)
Purpose To read Model 7700 calibration lock status.
Format :cal:prot:card1:lock?
Response 0 Calibration locked.
1 Calibration unlocked.
Description The :LOCK? query requests status from the Model 2701 on Model 7700
calibration locked/unlocked state. Calibration must be enabled sending the
:CODE command before calibration can be performed.
Example :CAL:PROT:CARD1:LOCK? Request card 1 cal lock state.
:RCOunt
(:CALibration:PROTected:CARD1:RCOunt)
Purpose To reset card calibration count to 0.
Format :cal:prot:card1:rco
Parameter None
Description The :RCOunt command resets the card calibration count reported by
:COUNt? to 0.
Example :CAL:PROT:CARD1:RCO Rest card calibration count.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals