B-8 Calibration Reference Model 2701 Service Manual
:DATA?
(:CALibration:PROTected:DATA?)
Purpose To read calibration constants data.
Format :cal:prot:data?
Response Floating-point, ASCII values separated by commas.
Description The :DATA? query requests all the calibration constants. Values are returned
in one string of ASCII floating-point values delimited by commands.
Example :CAL:PROT:DATA? Request cal constants.