EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 Service Manual

Keithley 2701
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
B-8 Calibration Reference Model 2701 Service Manual
:DATA?
(:CALibration:PROTected:DATA?)
Purpose To read calibration constants data.
Format :cal:prot:data?
Response Floating-point, ASCII values separated by commas.
Description The :DATA? query requests all the calibration constants. Values are returned
in one string of ASCII floating-point values delimited by commands.
Example :CAL:PROT:DATA? Request cal constants.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals