Model 2701 Service Manual Performance Verification 1-5
Verification limits
The verification limits stated in this section have been calculated using only the Model 2701
one-year accuracy specifications, and they do not include test equipment uncertainty. If a
particular measurement falls slightly outside the allowable range, recalculate new limits based
on both Model 2701 specifications and pertinent calibration equipment specifications.
Example reading limit calculation
The following is an example of how reading limits have been calculated. Assume you are
testing the 10V DC range using a 10V input value. Using the Model 2701 one-year accuracy
specification for 10V DC of ± (30ppm of reading + 5ppm of range), the calculated limits are:
Reading limits = 10V ± [(10V
×
30ppm) + (10V
×
5ppm)]
Reading limits = 10V ± (0.0003 + 0.00005)
Reading limits = 10V ± 0.00035V
Reading limits = 9.99965V to 10.00035V
Calculating resistance reading limits
Resistance reading limits must be recalculated based on the actual calibration resistance
values supplied by the equipment manufacturer. Calculations are performed in the same
manner as shown in the preceding example, except, of course, that you should use the actual
calibration resistance values instead of the nominal values when performing your calculations.
For example, assume that you are testing the 10k
range using an actual 10.03k
calibration resistance value. Using Model 2701 one-year 10k
range accuracy of ± (100ppm of
reading + 6ppm of range), the calculated reading limits are:
Reading limits = 10.03k
± [(10.03k
×
100ppm) + (10k
×
6ppm)]
Reading limits = 10.02894k
to 10.03106k
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory defaults
as follows:
1. Press
SHIFT
and then
SETUP
. The instrument will display the following prompt:
RESTORE: FACT.
2. Using either range key, select FACT, then restore the factory default conditions by
pressing
ENTER
.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals