EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 Service Manual

Keithley 2701
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Model 2701 Service Manual Specifications A-9
Optimizing measurement accuracy
The configurations listed below assume that the multimeter has had factory setups restored.
DC voltage, DC current, and resistance:
Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range.
Use REL on DC voltage and 2-wire resistance measurements.
Use 4-wire resistance measurements for best accuracy.
AC voltage and AC current:
Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range.
Temperature:
Select 6½ digits, 10 PLC, filter ON (up to 100 readings).
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups restored.
DC voltage, DC current, and resistance:
Select 3½ digits, 0.01 PLC, filter OFF, fixed range.
AC voltage and AC current:
Select 3½ digits, 0.01 PLC, filter OFF, fixed range.
Temperature:
Select 3½ digits, 0.01 PLC, filter OFF.
For all functions, turn off the display and autozero, and set the trigger delay to zero. Use the
:SAMPle:COUNt and READ? bus commands.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals