EasyManuals Logo

Keithley 2701 Service Manual

Keithley 2701
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
B-16 Calibration Reference Model 2701 Service Manual
Manufacturing calibration commands
Three calibration steps are only performed at the factory or when the unit has been repaired:
:CALibration:PROTected:AC:STEP14 1V AC at 3Hz
:CALibration:PROTected:AC:STEP15 1V AC at 1kHz
:CALibration:PROTected:DC:STEP0 Rear scanner terminal short circuit
:AC:STEP<14|15>
(CALibration:PROTected:AC:STEP<14|15>)
Purpose To program individual AC manufacturing calibration steps.
Format :cal:prot:ac:step14 <Cal_voltage>
:cal:prot:ac:step15 <Cal_frequency>
Parameter <Cal_voltage> = 1 [1V nominal]
<Cal_frequency = 1E3 [1kHz nominal]
Description The :AC:STEP14 and :AC:STEP15 commands program the two
manufacturing AC calibration steps. The appropriate signal must be
connected to the instrument when programming each step, as summarized
by the parameters listed above.
Example :CAL:PROT:AC:STEP14 1 Program AC step 14.
:CAL:PROT:AC:STEP15 1E3 Program AC step 15.
:DC:STEP0
(:CALibration:PROTected:DC:STEP0)
Purpose To perform rear scanner terminal short-circuit calibration.
Format :cal:prot:dc:step0
Parameter None
Description :STEP0 performs the rear scanner terminal short-circuit calibration step in
the manufacturing calibration procedure. Install an extender card with low-
thermal shorts on the output terminals installed in Slot 1 and select the rear
scanner inputs before sending this command.
Example :CAL:PROT:DC:STEP0 Perform rear short-circuit calibration.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals