Model 2701 Service Manual Calibration Reference B-5
:INIT
(:CALibration:PROTected:INITiate)
Purpose To initiate comprehensive and factory calibration procedures.
Format :cal:prot:init
Parameter None
Description The :INIT command enables Model 2701 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the :CODE command, but before sending any other calibration
command.
NOTE The
:INIT
command should be sent only once before performing either DC, AC, or
factory calibration. Do not send
:INIT
before each calibration step.
Example :CAL:PROT:INIT Initiate calibration.
:LOCK
(:CALibration:PROTected:LOCK)
Purpose To lock out comprehensive or manufacturing calibration.
Format :cal:prot:lock
Parameter None
Description The :LOCK command allows you to lock out both comprehensive and
manufacturing calibration after completing those procedures. Thus, :LOCK
performs the opposite of enabling calibration with the :CODE command.
NOTE To unlock comprehensive calibration, send the
:CODE
command. To unlock
manufacturing calibration, hold in the OPEN key while turning on the power.
Example :CAL:PROT:LOCK Lock out calibration.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals