13-18 Return to Section Topics 6482-901-01 Rev. A / August 2012
Section 13: Status Structure Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Programming example: Read error queue
The following command reads the error queue:
STAT:QUE?
Table 13-9
Error queue commands
Command Description Comment
STATus
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
SYSTem
:ERRor
[:NEXT]?
:ALL?
:COUNt?
:CODE
[:NEXT]?
:ALL?
:CLEar
STATus subsystem:
Read error queue:
Read and clear oldest error and status (code and message).
Specify error and status messages for error queue.
Read the enabled messages.
Specify messages not to be placed in queue.
Read the disabled messages.
Clear messages from error queue.
SYSTem subsystem:
Read error queue:
Read and clear oldest error and status (code and message).
Read and clear all errors and status (code and message).
Read the number of messages in queue.
Code numbers only:
Read and clear oldest error and status (code only).
Read and clear all errors and status (codes only).
Clear messages from error queue.
Note 1
Note 2
Note 2
Note 1
Notes:
1. Power-up and *CLS empty the error queue. STATus:PRESet has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.