16-52 Return to Section Topics 6482-901-01 Rev. A / August 2012
Section 16: SCPI Command Reference Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
DISable <list>
:STATus:QUEue:DISable <list> Disable messages for error queue
Parameters <list> = (numlist)
where numlist is a specified list of messages that you wish to disable for the error
queue.
Query :DISable? Query list of disabled messages
Description On power-up, all error messages are enabled and will go into the error queue as
they occur. Status messages are not enabled and will not go into the queue. This
command is used to specify which messages you want disabled. Disabled
messages are prevented from going into the error queue.
SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in Table 16-8.
Default conditions
PRESet
:SYSTem:PRESet Return to :SYSTem:PRESet defaults
Description This command returns the instrument to states optimized for front-panel
operation. :SYSTem:PRESet defaults are listed in the SCPI tables (Table 16-1
through Table 16-10).
POSetup
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet defaults
SAV0 Power-up to setup stored at memory location 0
SAV1 Power-up to setup stored at memory location 1
SAV2 Power-up to setup stored at memory location 2
SAV3 Power-up to setup stored at memory location 3
SAV4 Power-up to setup stored at memory location 4
Query :POSetup? Query power-on setup
Description This command is used to select the power-on defaults. With RST selected, the
instrument powers up to the *RST default conditions. With PRES selected, the
instrument powers up to the :SYStem:PRESet default conditions. Default
conditions are listed in the SCPI tables (Table 16-1 through Table 16-10).
With the SAV0-4 parameters specified, the instrument powers-on to the setup that
is saved in the specified location using the *SAV command.