17-16 Return to Section Topics 6482-901-01 Rev. A / August 2012
Section 17: Calibration Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Changing the password
The default password (KI006482) may be changed from the front-panel or over a remote interface,
as discussed below.
Changing the password from the front-panel
Follow these steps to change the password from the front-panel:
1. Press the MENU key, select CAL, and then press the ENTER key. The instrument will display
the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
2. Select UNLOCK then enter the password. Default: 006482.
3. Select CHANGE-PASSWORD and press the ENTER key. The instrument will display the
following:
New Pwd: 006482
Use , , , , ENTER or EXIT.
4. Using the EDIT keys, enter the new password on the display.
5. When finished entering the new password, press the ENTER key to store the new password.
Changing the password through a remote interface
To change the calibration password through a remote interface, first send the present password,
and then send the new password. For example, the following command sequence changes the
password from the ‘KI006482’ remote default to ‘KI_CAL’:
:CAL:PROT:CODE 'KI006482'
:CAL:PROT:CODE 'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTE If you change the first two characters of the password to something other than “KI”, you
will not be able to unlock calibration from the front-panel.
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the CAL pads,
which are located on the display board. Doing so will also reset the password to the factory default
(006482, front-panel; KI006482, remote).