6482-901-01 Rev. A / August 2012 Return to Section Topics F-5
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual Appendix F: Example Programs
Limit test program
The program listing below performs a limits test as covered in Section 10. This program sets up
the following operating modes:
Source and measure channel: channel 1
Source voltage: 1 V
Source delay: 100 ms
Limit 3 upper value: 10 mA
Limit 3 lower value: 4 mA
Limit 4 upper value: 8 mA
Limit 4 lower value: 6 mA
' Limit Test Program
*RST ' Restore GPIB defaults.
:SENS1:CURR:RANG:AUTO ON ' Channel 1 autorange.
:SOUR1:VOLT:TRIG 10 ' Source 1 V when triggered.
:SOUR1:DEL 0.1 ' 0.1 s source delay.
:CALC7:FEED CALC3 ' Use Channel 1 data for limit tests.
:CALC7:LIM3:UPP 10e-3 ' Limit 3 upper limit = 10 mA.
:CALC7:LIM3:LOW 4e-3 ' Limit 3 lower limit = 4 mA.
:CALC7:LIM4:UPP 8e-3 ' Limit 4 upper limit = 8 mA.
:CALC7:LIM4:LOW 6e-3 ' Limit 4 lower limit = 6 mA.
:CALC7:LIM1:STAT 0 ' Turn off Limit 1.
:CALC7:LIM3:STAT 1 ' Turn on Limit 3.
:CALC7:LIM4:STAT 1 ' Turn on Limit 4.
:OUTP1 ON ' Turn on channel 1 output.
:INIT ' Trigger readings.
:OUTP1 OFF ' Turn off channel 1 output.
:CALC7:LIM3:FAIL? ' Request Limit 3 fail status.
' Read back result
:CALC7:LIM4:FAIL? ' Request Limit 4 fail status.
' Read back result