6.6.1 Switching on/off
Start-up Press the ON/OFF button for approx. 0.5 sec.
The balance will carry out a self-test The balance is ready for
weighing when the weight display appears.
Switching Off Press the ON/OFF button for approx. 0.5 sec.
-OFF- appears briefly before the display disappears.
6.6.2 Stability display
The appearance of the stability symbol [ ] on the display indicates that the
weighing plate is in a stable state. If the status is instable the [ ] display
disappears.
6.6.3 Balance zero display
Should the balance not display exactly zero despite empty scale pan, press the
button. The balance starts with resetting to zero [ a ] .
7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Procedure when adjusting:
Calibration should be carried out by using a recommended calibration weight.
(See chpt. 1 “Technical specifications“):
Observe stable environmental conditions. A warm-up time of 2 hours is necessary for
stabilisation.