PLE-BA-e-0812
30
12.4 AUTO-OFF - Automatic switching-off
If the AUTO-OFF function is activated, the balance switches off automatically to save
the rechargeable batteries after 5 minutes without load change.
Function
Adjustment
Mains power
supply
Battery power
supply
t1 = no deactivated deactivated
t1 = YES activated activated
t1 = Auto deactivated activated
Ö Press the + button simultaneously, "P1 rEAd" will appear
⇒ Press the button as often as necessary
until the required setting appears:
⇒
⇒ Press repeatedly.
⇒
⇒ Press repeatedly.
⇒
Ö Press the button repeatedly until the "SAVE"? inquiry appears.
Confirm the inquiry by pressing the button to save the changes and to
return to Weighing mode.
-or-
press the button to return to Weighing mode without saving.