11.3 Auto-Zero – Automatic Zeroing
This function is used to tare small variations in weight automatically.
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability
compensation“ in the balance. (e.g. slow flow of liquids from a container placed on
the balance, evaporating processes).
When apportioning involves small variations of weight, it is advisable to switch off this
function.
Call up menu:
Ö Press the + button simultaneously, "P1 rEAd" will appear
Press repeatedly until the wanted
setting appears (no = off, yes = on)
Ö Press the button repeatedly until the "SAVE"? inquiry appears.
Confirm the inquiry by pressing the
button to save the changes and to
return to Weighing mode.
-or-
press the
button to return to Weighing mode without saving.