14.3 Communication protocol / remote control commands
Instruction: Remote control instructions
Z
Set weight display at zero
T
Taring
S
Send stable weighing value in standard weighing unit
SI
Send weighing value immediately in standard weighing unit
SU
Send stable weighing value in current weighing value
SUI
Send weighing value immediately in current weighing unit
C1
Turn on continuous transmission in standard weighing unit
C0
Turn off continuous transmission in standard weighing unit
CU1
Turn on continuous transmission in current weighing unit
CO1
Turn off continuous transmission in current weighing unit
PC
Send all implemented instructions
Complete each instruction with CR LF.
14.3.1 Response messages from balance
Instruction: Response messages from balance
XX_
Instruction:
XX_A CR LF
Instruction accepted; will be executed
XX_D CR LF
Instruction complete (appears after XX_A only)
XX_I CR LF
Instruction received; impossible to carry out
XX _ ^ CR LF
Instruction received but time overflow error occurred
XX _ v CR LF
Instruction received, but insufficient load
XX _ E CR LF
Error during execution, timeout for stable weighing value
exceeded