PLE-BA-e-0812
56
Data record format:
1-3 4 5 6 7-15 16 17 18 19
Instruction:
Stability indicator
Blank
Signs
Weight
Blank
Unit
CR
LF
Stability indicator [blank space], if stable
[?] if not stable
[^] if overload
[v] if underload
Presign: [blank space] if positive
[-] if negative
Weight: 9 signs, right justified
Unit: 3 signs, left justified
Instruction: 3 signs, left justified
14.4 Manual output
The user can start output manually by pressing the PRINT-key (for settings see
chapter 14.2.2).
Data record format:
1 2 3 4 - 12 13 14 - 16 17 18
Stability indicator
Blank
Signs
Weight
Blank
Unit
CR
LF
Stability indicator [blank space], if stable
[?] if not stable
[^] if overload
[v] if underload
Presign: [blank space] if positive
[-] if negative
Weight: 9 signs, right justified
Unit: 3 signs, left justified