FETCh:PERiod? [<first_pt>, [<last_pt>]]
Performs post-processing calculation on the specified range of readings and returns period for single channel DMM
digitizing.
For valid post processed measurement, please ensure data captured is at least 1 cycle.
This SCPI is supported on DMM Digitize data acquisition mode only.
Parameter Typical return
<first_pt> = An integer indicating the first data point of the range of readings to perform
post processing calculation to obtain the period.
Default: 1
<last_pt> = An integer indicating the last data point of the range of readings to perform
post processing calculation to obtain the period.
Default: 256
(none)
Returning the period of the range of readings from data number 2 to 500
FETCh:PER? 2,500
Typical response: +2.00908184E-02
FETCh:PWIDTh? [<first_pt>, [<last_pt>]]
Performs post-processing calculation on the specified range of readings and returns positive width for single
channel DMM digitizing.
For valid post processed measurement, please ensure data captured is at least 1 cycle.
This SCPI is supported on DMM Digitize data acquisition mode only.
Parameter Typical return
<first_pt> = An integer indicating the first data point of the range of readings to perform
post processing calculation to obtain the positive width.
Default: 1
<last_pt> = An integer indicating the last data point of the range of readings to perform
post processing calculation to obtain the positive width.
Default: 256
(none)
Returning the positive width of the range of readings from data number 2 to 500
FETCh:PWIDTh? 2,500
Typical response: +9.73982085E-03
122
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming