FETCh3 Subsystem
These SCPIs are supported on the DAQM909A Digitizer module only.
The data of incomplete records cannot be retrieved.
Example #1: The digitizer channel is configured as such - Arm Count = 1, Sample Count = 1,000,000 and the
digitizing operation is initiated. If you abort the operation before 1,000,000 samples have been acquired, all
acquired data in the 1st Record prior to the abort cannot be retrieved.
Example #2: The digitizer channel is configured as such - Arm Count = 3, Sample Count = 1,000,000 and the
digitizing operation is initiated. If you abort the operation during the 3rd Record before 1,000,000 samples
have been acquired, all acquired data in the 3rd Record prior to the abort cannot be retrieved. However the
acquired data in the 1st and 2nd Record is still retrievable.
Command Summary
FETCh3
FETCh3:DATA
FETCh3:{DCYCle | DUTYcycle}
FETCh3:FFTRansform:MAGNitude
FETCh3:FFTRansform:PHASe
FETCh3:FREQuency
FETCh3:{FTIMe|FALL:TIME}
FETCh3:NWIDth
FETCh3:PERiod
FETCh3:POWer
FETCh3:PWIDth
FETCh3:RMS?
FETCh3:{RTIMe|RISE:TIME}
FETCh3:STATistics
FETCh3:THDNoise
FETCh3:THDistortion
Keysight DAQ970A/DAQ973A Programming Guide 125
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies DAQ970A and is the answer not in the manual?

Keysight Technologies DAQ970A Specifications

General IconGeneral
BrandKeysight Technologies
ModelDAQ970A
CategoryMultiplexer
LanguageEnglish