ACQuire Subsystem
ACQuire commands provide the way to put the instrument into DMM Digitize data acquisition mode. When you
execute these commands, the instrument uses default values for the requested measurement configuration.
But unlike CONFigure commands, ACQuire commands sets the NPLC to minimum, disables both autoranging
and autozero. ACQuire commands also let you select the sample count and sample time for DMM Digitize
mode.
The ACQuire command does not place the instrument in the "wait-for-trigger" state. Use theINITiate or READ?
command in conjunction with ACQuire to place the instrument in the "wait-for-trigger" state.
To exit from DMM Digitize data acquisition mode, you may use the command ROUTe:SCAN (@).
Command Summary
ACQuire?
ACQuire:CURRent[:DC]
ACQuire:VOLTage[:DC]
ACQuire:RESistance
ACQuire:FRESistance
ACQuire:STRain:DIRect
ACQuire:STRain:FDIRect
ACQuire:STRain:QUARter
ACQuire:STRain:HALF:BENDing
ACQuire:STRain:FULL:BENDing
ACQuire:STRain:HALF:POISson
ACQuire:STRain:FULL:POISson
ACQuire:STRain:FULL:BENDing:POISson
ACQuire:TEMPerature:FRTD
ACQuire:TEMPerature:RTD
ACQuire:TEMPerature:TCouple
Keysight DAQ970A/DAQ973A Programming Guide 35
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies DAQ970A and is the answer not in the manual?

Keysight Technologies DAQ970A Specifications

General IconGeneral
BrandKeysight Technologies
ModelDAQ970A
CategoryMultiplexer
LanguageEnglish