ACQuire:TEMPerature:{FRTD|RTD} [{<r0> | MIN | MAX | DEF}, [{<samp_count> | MIN | MAX | DEF},
[{<samp_time> | MIN | MAX | DEF},]]] (@<channel>)
Configures a single channel for 2-wire or 4-wire RTD temperature DMM digitizing, but they do not initiate the
digitizing. This command redefines the scan list.
Parameter Typical return
<r0>
MIN: 100 Ω
MAX: 1000 Ω
DEF: 100 Ω
<samp_count>
MIN: 1
MAX: 50,331,648
DEF: 1
<samp_time>
MIN: 20 μs
MAX: 3600 s
DEF: 1 s
(none)
Configures a default 2-wire RTD temperature measurement on Channel 101:
ACQ:TEMP:RTD DEF,DEF,DEF,(@101)
44
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming