ACQuire:TEMPerature:TCouple [{B | E | J | K | N | R | S | T | DEF}, [{<samp_count> | MIN | MAX | DEF},
[{<samp_time> | MIN | MAX | DEF},]]] (@<channel>)
Configures a single channel for thermocouple temperature DMM digitizing, but they do not initiate the digitizing.
This command redefines the scan list.
Parameter Typical return
<type> = {B | E | J | K | N | R | S | T | DEF}
DEF: J
<samp_count>
MIN: 1
MAX: 50,331,648
DEF: 1
<samp_time>
MIN: 20 μs
MAX: 3600 s
DEF: 1 s
(none)
Configures a default thermocouple temperature measurement on Channel 101:
ACQ:TEMP:TC DEF,DEF,DEF,(@101)