ABORt
Aborts a measurement in progress from a scan, returning the instrument to the trigger idle state.
Parameter Typical return
(none) (none)
Aborts the measurement in progress: ABOR
– If a scan is in progress when the command is received, the scan will not be completed and you cannot resume
the scan from where it left off. Note that if you initiate a new scan, all readings are cleared from reading memory.
– The Factory Reset (*RST command) will abort a measurement, clear the scan list, and set all measurement para-
meters to the factory configuration. The Instrument Preset (SYSTem:PRESet command) also aborts a meas-
urement but will not clear the scan list.
ABORt3
Aborts module digitizing for all the channels that have been initiated, returning them to the trigger idle state.
This SCPI is only supported on the DAQM909A Digitizer module.
Parameter Typical return
(none) (none)
Aborts the digitizing operation in progress: ABOR3