ACQuire:{RESistance|FRESistance} [{<range> | MIN | MAX | DEF},
[{<samp_count> | MIN | MAX | DEF}, [{<samp_time> | MIN | MAX | DEF},]]] (@<channel>)
Configures a single channel for 2-wire or 4-wire resistance DMM digitizing, but they do not initiate the digitizing.
This command redefines the scan list.
RESistance and FRESistance measurements are only supported on DAQM900A, DAQM901A, DAQM902A, and
DAQM908A multiplexer modules.
Parameter Typical return
<range> = {100 Ω | 1 | 10 | 100 | 1 MΩ | 10 | 100 | 1GΩ}
MIN: 100 Ω
MAX: 1GΩ
DEF: 1
<samp_count>
MIN: 1
MAX: 50,331,648
DEF: 1
<samp_time>
MIN: 20 μs
MAX: 3600 s
DEF: 1 s
(none)
Configures a default resistance measurement on Channel 101:
ACQ:RES DEF,DEF,DEF,(@101)