ACQuire:VOLTage[:DC] [{<range> | MIN | MAX | DEF}, [{<samp_count> | MIN | MAX | DEF},
[{<samp_time> | MIN | MAX | DEF},]]](@<channel>)
Configures a single channel for DC voltage DMM digitizing, but they do not initiate the digitizing. This command
redefines the scan list.
Parameter Typical return
<range> = {100 mV | 1 V | 10 V | 100 V | 300 V}
MIN: 100 mV
MAX: 300 V
DEF: 300 V
<samp_count>
MIN: 1
MAX: 50,331,648
DEF: 1
<samp_time>
MIN: 20 μs
MAX: 3600 s
DEF: 1 s
(none)
Configures a default voltage measurement on Channel 101:
ACQ:VOLT:DC DEF,DEF,DEF,(@101)