EasyManuals Logo

Keysight Technologies DAQ970A User Manual

Keysight Technologies DAQ970A
393 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
FETCh:STATistics? [<first_pt>, [<last_pt>]]
Performs post-processing calculation on the specified range of readings and returns statistics for single channel
DMM digitizing.
This SCPI is supported on DMM Digitize data acquisition mode only.
Parameter Typical return
<first_pt> = An integer indicating the first data point of the range of readings to perform
post processing calculation to obtain the statistics.
Default: 1
<last_pt> = An integer indicating the last data point of the range of readings to perform
post processing calculation to obtain the statistics.
Default: 256
(none)
Returning the statistics of the range of readings from data number 2 to 500.
The returned statistics value is in this order: average, standard deviation, minimum, and maximum.
FETC:STAT? 2,500
Typical response: -6.17764966E-04,+5.13837757E-02,-7.68353078E-02,+7.65340847E-02
FETCh:{THDNoise|THDistortion}? [<first_pt>, [<length>]]
Performs post-processing calculation on the specified range of readings and returns THD noise/distortion for single
channel DMM digitizing.
The minimum data points to perform the post-processing calculation is 256 data points. For valid post
processed measurement, please ensure data captured is at least 1 cycle.
The <length> will be round down to the lower 2
n
number, if the number enter is not 2
n
.
This SCPI is supported on DMM Digitize data acquisition mode only.
Parameter Typical return
<first_pt> = An integer indicating the first data point of the range of readings to perform
post processing calculation to obtain the statistics.
Default: 1
<length> = An 2
n
integer (256, 512,1024, 2048, 4096, …., 65536) indicating the number of
data points to perform post processing calculation to obtain the THD noise/ distortion.
Default: 256
(none)
Returns the THD noise of 256 readings starting from the second data point
FETCh:THDNoise? 2,256
Typical response: -3.16766701E+01
124
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies DAQ970A and is the answer not in the manual?

Keysight Technologies DAQ970A Specifications

General IconGeneral
BrandKeysight Technologies
ModelDAQ970A
CategoryMultiplexer
LanguageEnglish