EasyManua.ls Logo

Keysight Technologies DAQ970A - Page 135

Keysight Technologies DAQ970A
393 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FETCh3:RMS? [<record #>, [<first_pt>, [<last_pt>,]]] (@<channel>)
Performs post-processing calculation on the specified range of readings and returns RMS on the specified record
for a channel.
This SCPI is supported on the DAQM909A Digitizer module only.
Parameter Typical return
<record #> = An integer indicating the specified record for a channel.
Default: 1
<first_pt> = An integer indicating the first data point of the range of readings to perform
post processing calculation to obtain the RMS.
Default: 1
<last_pt> = An integer indicating the last data point of the range of readings to perform
post processing calculation to obtain the RMS.
Default: 256
<channel> = Channel number
(none)
Returning the RMS of the range of readings from data number 2 to 500 in the first record of Digitizer channel 201.
FETCh3:RMS? 1,2,500,(@201)
Typical response: +5.13359793E-02
Keysight DAQ970A/DAQ973A Programming Guide 135
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A