[SENSe:]TOTalize:STARt [(@<ch_list>)]
Starts totalizing on the specified totalizer channels.
Parameter Typical return
(none) (none)
Starts totalizing on totalizer channel:TOT:STAR:IMM
[SENSe:]TOTalize:STOP [(@<ch_list>)]
Stops totalizing on the specified totalizer channels.
Parameter Typical return
(none) (none)
Stops totalizing on totalizer channel:TOT:STOP:IMM
[SENSe:]TOTalize:TYPE {READ | RRESet} [, (@<ch_list>)]
[SENSe:]TOTalize:TYPE? [(@<ch_list>)]
Enables or disables an automatic reset of the count on the specified totalizer channels.
READ: Read the totalizer during a scan without resetting the count.
PRESet: Read the totalizer during a scan and reset the count to 0 after it is read.
Parameter Typical return
{READ|RRESet}
Default: READ
READ or PRES
Configures totalizer channel to be read without resetting its count: TOT:TYPEREAD
The maximum count is 67,108,863 (2
26
- 1). The count rolls over to 0 after reaching the maximum allowed
value.