SENSe3:STRain:VOLTage:RANGe[:UPPer] {<range> | MIN | MAX | DEF} , (@<ch_list>)
SENSe3:STRain:VOLTage:RANGe[:UPPer]? {(@<ch_list>) | MIN | MAX | DEF}
Sets a fixed voltage range for strain measurements on the specified channels.
This SCPI is supported on the DAQM909A Digitizer module only.
Parameter Typical return
<range> = {300 mV | 1V | 3V | 18V}
MIN: 300 mV
MAX: 18 V
DEF: 1 V
+1.80000000E+01
Sets the maximum voltage range for strain measurement on Digitizer channel 201
SENS3:STR:VOLT:RANG MAX,(@201)
SENSe3:VOLTage[:DC]:RANGe[:UPPer] {<range> | MIN | MAX | DEF}, (@<ch_list>)
SENSe3:VOLTage[:DC]:RANGe[:UPPer]? {(@<ch_list>) | MIN | MAX | DEF}
Sets a fixed voltage range for DC voltage measurements on the specified channels.
This SCPI is supported on the DAQM909A Digitizer module only.
Parameter Typical return
<range>: {300 mV | 1 V | 3 V | 18 V}.
MIN: 300 mV
MAX: 18 V
DEF: 1 V
+1.80000000E+01
Sets a 18 V voltage range for DC voltage measurement on Digitizer channel 201
SENS3:VOLT:DC:RANG MAX,(@201)
Keysight DAQ970A/DAQ973A Programming Guide 287
2 SCPI Programming