Module related self-test errors
Error code Error statement
671 Self-test failed; communication with module slot #
672 Self-test failed; service request of module slot #
673 Self-test failed; GP0 signal of module slot #
Self-test failed; GP1 signal of module slot #
674 Self-test failed; temperature sensor of module slot #
Self-test failed; communication with temperature sensor of module slot #
675 Self-test failed; DAC autocal output voltage offset of module slot #, channel #
676 Self-test failed; DAC autocal output voltage positive gain of module slot #, channel #
677 Self-test failed; DAC autocal output voltage negative gain of module slot #, channel #
678 Self-test failed; DAC autocal output current offset of module slot #, channel #
679 Self-test failed; DAC autocal output current positive gain of module slot #, channel #
680 Self-test failed; DAC autocal output current negative gain of module slot #, channel #
681 Self-test failed; DAC autocal sense voltage offset of module slot #, channel #
682 Self-test failed; DAC autocal sense voltage gain of module slot #, channel #
683 Self-test failed; DAC autocal sense current offset of module slot #, channel #
684 Self-test failed; DAC autocal sense current gain of module slot #, channel #
685 Self-test failed; nonvolatile RAM checksum of NvRelayFactory of module in slot #
Self-test failed; nonvolatile RAM checksum of NvRelayOffset of module in slot #
Self-test failed; fail to get nonvolatile RAM checksum of module in slot #
686 Self-test failed; fail to get firmware image checksum of module in slot #
688 Self-test failed; communication with module FPGA slot #
689 Self-test failed; communication with module ADC slot #
690 Self-test failed; module High Speed Link slot #
691 Self-test failed; Bist Relays test of module slot #, 1VSupply fail
Self-test failed; Bist Relays test of module slot #, 1.1VSupply fail
Self-test failed; Bist Relays test of module slot #, 2.5VSupply fail
Self-test failed; Bist Relays test of module slot #, 3.66VSupply fail
Self-test failed; Bist Relays test of module slot #, 4.67VSupply fail
Self-test failed; Bist Relays test of module slot #, 5VSupply fail
Self-test failed; Bist Relays test of module slot #, N4VAutoCal fail
Self-test failed; Bist Relays test of module slot #, P4VAutoCal fail
Self-test failed; Bist Relays test of module slot #, 1VShunt fail
Self-test failed; Bist Relays test of module slot #, 1.1VShunt fail
Self-test failed; Bist Relays test of module slot #, 2.5VShunt fail
Self-test failed; Bist Relays test of module slot #, 3.66VShunt fail
Self-test failed; Bist Relays test of module slot #, 4.67VShunt fail
Self-test failed; Bist Relays test of module slot #, 5VBackplane fail
Self-test failed; Bist Relays test of module slot #, 3.3VSupply fail
Self-test failed; Bist Relays test of module slot #, 1VAutoCal fail