CALibration Subsystem
For a detailed discussion on the calibration procedures, see Calibration Adjustment Procedures.
CALibration:COUNt?
Returns the number of times the instrument has saved calibration data as a signed whole number, for
example +21. Your instrument was calibrated at the factory; read and record the initial count when you
receive the instrument.
CALibration:CURRent[:DATA][:HIGH] <value>
Enters an output current value (in amps) that you obtained by reading a DMM. You must first select a
calibration level (CAL:CURR:LEV) for the value being entered. This command can only be used when
calibration is unsecured and output is ON.
CALibration:CURRent[:DATA]:LOW <value>
Enters an output current value (in amps) that you obtained by reading a DMM. This is similar to
CALibration:CURRent[:DATA][:HIGH], but it is only used for small current calibration. This command can
only be used when calibration is unsecured and output is ON.
CALibration:CURRent:LEVel[:HIGH] MINimum|MAXimum
Initiates calibration for a specific current level. The MINimumcalibration must be performed before the
MAXimum, and you must unsecure this instrument to use this command.
CALibration:CURRent:LEVel:LOW MINimum|MAXimum
Initiates small current calibration for a specific current level. The MINimumcalibration must be
performed before the MAXimum, and you must unsecure this instrument to use this command.