SYSTem:RWLock
Places the power supply in the remote mode. This command is the same as SYSTem:REMote, except
that all front-panel keys, excluding the [Lock/Unlock] key are disabled. You can unlock the front panel
by holding the [Lock/Unlock] key. This switches the instrument to local mode at the same time. The
front panel will be unlocked when the instrument is switched to remote or local mode.
SYSTem:SECurity:IMMediate
Clears and sanitizes all user memory (states, LAN information, and so on), usually to prepare the
instrument for removal from a secure area. Instrument identification data (instrument firmware, model
number, serial number, MAC address and calibration data) is not erased. This procedure is not
recommended for routine use because of the possibility of unintended data loss.
SYSTem:VERSion?
Returns the SCPI version on which the instrument's command set is based, which is 2005.0.