OUTPut Subsystem
OUTPut[:STATe] ON |1 | OFF |0
OUTPut[:STATe]?
Enables or disables the instrument's output.The query returns 0 (OFF) or 1 (ON). At *RST, the output
state is off.
OUTPut:PROTection:CLEar
Clears the latch that disables the output due to an overvoltage or overcurrent condition. You must clear
the conditions that cause the fault before executing this command. You can then restore the output to
the state that existed before the fault condition occurred.
OUTPut:PON:STATe
RST|RCL0|RCL1|RCL2|RCL3|RCL4|RCL5|RCL6|RCL7|RCL8|RCL9|
OUTPut:PON:STATe?
Specifies whether the instrument's power-on state is the *RST state (default) or the state stored in one
of ten memory locations.