Save or recall the instrument's state
The instrument allows you to save and recall the instrument state in one of 10 memory locations, labeled
0through9. This allows you to quickly configure the instrument for commonly used applications.
An instrument state includes the instrument's output enable status, voltage and current settings, and
OCP and OCV settings. It does not include error codes or LAN configuration information.
To save or recall the instrument state
1. Press [Menu].
2. Turn the knob to Store (or Recall) and press the knob.
3. Turn the knob to an option from Store 0 to Store 9 (or Recall 0 to Recall 9)and press the knob to
store or recall the present state of the instrument in the specified memory location.
4.
The instrument will briefly display DONE.
Configure the Power-on State
The instrument allows you to recall the power-on state as either the default (*RST) state or the state
stored in one of 10 memory locations (0through9). To configure the power-on state:
1. Press [Menu].
2. Turn the knob to Power-on and press the knob.
3.
Turn the knob to Default, or a state from Recall 0 to Recall 9 and press the knob to make your selec-
tion.
4.
The instrument will briefly display CHANGED SAVED if the recall state is different from the previous
state, and NOCHANGE if the same state was selected.