APPLy Subsystem
APPLy <voltage>| DEFault | MINimum | MAXimum[,<current>| DEFault |
MINimum | MAXimum]
APPLy?
Specifies and changes the output voltage and current in one command.
The DEFault, MINimum, and MAXimum values for each mode are shown in Programming Ranges.
The query returns a quoted string with the voltage setting before the current setting.
Both values returned by the query have five decimal points: "25.00000,0.75000"