EasyManua.ls Logo

Keysight Technologies E36100 Series - Save the Calibration Data

Keysight Technologies E36100 Series
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step Front Panel SCPI
3 When the display shows second point calibration, connect the DMM current input
with the appropriate calibration resistor (see
Performance Verification >
Recommended test equipment
) for your instrument model to the power sup-
ply.
4 Measure output current with the DMM. CAL:CURR:LEV:LOW MAX
5 Enter the measured value, and press the knob. CAL:CURR:LOW <rdg>
6 Read DONE or FAIL on the display. (wait 25 seconds)
SYST:ERR?
Save the calibration data
To save calibration data, go to the ‘CAL Save’ menu to save the calibration data or enable the ‘Auto Save
feature. With Auto Save, calibration data will be saved when the user exits the calibration menu.
After completing the Voltage, Current and Small Current calibrations, save the calibration data before
exiting the calibration state, or simply exit the calibration state if Auto Save is on.
To save the CAL data: CAL:SAVE
To enable the CAL Auto Save: CAL:ASAVE ON
To exit CAL State: CAL:STAT OFF, 36102

Table of Contents

Related product manuals