CALibration:STATe <state>,<code>
CALibration:STATe?
Unsecures or secures the power supply with the calibration security code Of up to nine digits. The query
returns 1 (ON - unsecured for calibration) or 0 (OFF - secured against calibration).
The query only returns the state; it does not return the passcode.
The security code is nonvolatile, and does is not changed by power cycling or *RST.
The front panel limits the user to entering codes of up to six digits.
The default passcodes are 36102, 36103, 36104, 36105, and 36106.
Example: Unlock the instrument to enable calibration: CAL:STATe 1,36102
CALibration:STRing "<string>"
CALibration:STRing?
Record or read up to 40 characters of instrument calibration information.
You can store any information related to calibration, such as date or contact information.
The power supply must be unsecured before sending a calibration message.
You can read the string regardless of whether the supply is secured.
CALibration:VOLTage[:DATA] <value>
Enters an output voltage value (in volts) that you obtained by reading a DMM.
You must first select a calibration level (CAL:CURR:LEV) for the value being entered.
This command can only be used when calibration is unsecured and output is ON.
CALibration:VOLTage:LEVel MINimum|MAXimum
Initiates calibration for a specific voltage level. You must do the MINimum calibration before the
MAXimum, and this command can only be used when calibration is unsecured.