EasyManua.ls Logo

Keysight Technologies E36100 Series - Calibrate Voltage; Calibrate Current; Calibrate Small Current

Keysight Technologies E36100 Series
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrate voltage
Let the unit sit with output ON for one minute, then connect the DMM voltage input to the power
supply.
Step Front Panel SCPI
1 Turn the knob to Cal Voltage’ and press the knob. CAL:VOLT:LEV MIN
2 Measure the output voltage (low point) with the DMM.
3 Enter the measured value, and press the knob. CAL:VOLT <reading>
CAL:VOLT:LEV MAX
4 Measure the output voltage (high point) with the DMM.
5 Enter the measured value, and press the knob. CAL:VOLT <reading>
6 Read DONE or FAIL on the display. (wait 10 seconds)
SYST:ERR?
Calibrate current
Let the unit sit with output ON for one minute, then connect a current monitoring resistor across the
output terminals to be calibrated and then connect a DMM across the terminals of the monitoring
resistor.
Step Front Panel SCPI
1 Turn the knob to Cal Current and press the knob. CAL:CURR:LEV MIN
2 Measure the low output current with the DMM (DMM reading / shunt resistance).
3 Enter the measured value, and press the knob. CAL:CURR <reading>
CAL:CURR:LEV MAX
4 Measure the high output current with the DMM (DMM reading / shunt resistance).
5 Enter the measured value, and press the knob. CAL:CURR <reading>
6 Read DONE or FAIL on the display.
Calibrate small current
Allow the unit to sit with output ON for one minute before continuing.
Step Front Panel SCPI
1 Turn the knob to Cal Current and press the knob. CAL:CURR:LEV:LOW MIN
2 Wait approximately 40 seconds for the Cal in Prog message to leave the display.

Table of Contents

Related product manuals