EasyManua.ls Logo

Keysight Technologies E5260 Series

Keysight Technologies E5260 Series
350 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E5260/E5270 Programming Guide, Edition 4 4- 81
Command Reference
Command Reference
LSVM
The LSVM command selects the data output mode for the linear search
measurement.
Syntax LSVM mode
Parameters mode : Data output mode. Integer expression. 0 (initial setting) or 1.
0 : Returns Data_search only.
1 : Returns Data_search and Data_sense.
where
Data_search is the value forced by the search output channel set by the
LSV or LSI command.
Data_sense is the value measured by the search monitor channel set by
the LGI or LGV command.
For data output format, refer to “Data Output Format” on page 1-22.
Example
Statements
OUTPUT @E5270;"LSVM 1"
MCH
The MCH command selects the measurement channel for the data is displayed on
the front panel LCD.
Syntax MCH chnum
Parameters chnum : Measurement channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
Table 4-1 on page 4-9.
Example
Statements
OUTPUT @E5270;"MCH 1"

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Related product manuals