4- 80 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
LSV
The LSV command specifies and sets the voltage search source for the linear search
measurement. This command setting is cleared by the LSI command. After search
stops, the search channel forces the value specified by the LSM command.
Syntax LSV chnum,range,start,stop,step[,Icomp]
Parameters chnum : Search source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
Table 4-1 on page 4-9.
range : Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both start and stop values. Range
changing may cause 0
V output in a moment. For the limited auto
ranging, the instrument never uses the range less than the specified
range. See
Table 4-4 on page 4-13.
start, stop : Search start or stop voltage (in V). Numeric expression. See Table 4-6
on page 4-14. The start and stop parameters must have different values.
0 to ±100
0 to ±200 (for E5280B/E5290A)
step : Step voltage (in V). Numeric expression.
If start < stop, step must be positive, and if start > stop, step must be
negative. Maximum number of search steps is 1001.
Icomp: Current compliance value (in A). Numeric expression. See Table 4-6 on
page 4-14. If you do not specify Icomp, the previous value is set. Zero
amps (0 A) is not allowed for Icomp.
Example
Statements
OUTPUT @E5270;"LSV 1,0,0,20,.5,1E-6"