4- 100 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
RV
The RV command specifies the voltage measurement range or ranging type. In the
initial setting, the auto ranging is set. The range changing occurs immediately after
the trigger (that is, during the measurements). Voltage measurement channel can be
decided by the CMM command setting and the channel output mode (voltage or
current).
For the high speed spot measurement, use the TV/TTV command.
The range setting is cleared by the CL, CA, IN, *TST?, *RST or a device clear (HP
BASIC CLEAR) command.
Syntax RV chnum,range
Parameters chnum : Voltage measurement channel number. The value must be slot number
where the module has been installed. Integer expression. 1 to 8 (1 and 5
are not available for HPSMU). See
Table 4-1 on page 4-9.
range : Measurement range or ranging type. Integer expression. If you select
the fixed range, the instrument performs measurement by using the
specified range. For the auto or limited auto ranging, the measurement
range will be set to the minimum range that covers the measured
values. However, the instrument never uses the range less than the
specified range for the limited auto ranging. See
Table 4-2 on
page 4-10.
For the measurement mode that uses pulse source, if you select the auto
or limited auto ranging, the measurement channel uses the minimum
range that covers the compliance value or voltage output range.
Example
Statements
OUTPUT @E5270;"RV 2,-15"
OUTPUT @E5270;"RV 1,12"