Keysight E5260/E5270 Programming Guide, Edition 4 4- 111
Command Reference
Command Reference
TDI
Forces current and returns the time data from when the timer is cleared until output
is started. This command is effective for ASCII data output format. Refer to
“FMT”
on page 4-59.
Execution
Conditions
The CN command has been executed for the specified channel.
If the voltage compliance is greater than ±42 V, the interlock circuit must be shorted.
Syntax
TDI chnum,irange,current[,Vcomp[,comp_polarity[,vrange]]]
Parameters chnum : Source channel number. The value must be slot number where the
module has been installed. Integer expression. 1 to 8 (1 and 5 are not
available for HPSMU). See
Table 4-1 on page 4-9.
irange: Ranging type for current output. Integer expression. The output range
will be set to the minimum range that covers current value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
Table 4-5 on page 4-13.
current: Output current value (in A). Numeric expression. See Table 4-7 on
page 4-15.
0 to ±0.1, 0 to ±0.2 for E5291A, or 0 to ±1 for E5280B/E5290A
Vcomp: Voltage compliance value (in V). Numeric expression. See Table 4-7 on
page 4-15. If you do not specify this parameter, Vcomp is set to the
previous setting.
comp_
polarity:
Polarity of voltage compliance. Numeric expression.
0: Auto mode (default setting). The compliance polarity is
automatically set to the same polarity as current, regardless of the
specified Vcomp. If
current=0 A, the polarity is set to positive.
1: Manual mode. Uses the polarity of Vcomp you specified.
vrange: Voltage compliance ranging type. Integer expression. The compliance
range will be set to the minimum range that covers Vcomp value. For
the limited auto ranging, the instrument never uses the range less than
the specified range. See
Table 4-4 on page 4-13.
Example
Statements
OUTPUT @E5270;"TDI 1,0,1E-6"
ENTER @E5270 USING "#,5X,13D,X";Time

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals