4- 112 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
TDV
Forces voltage and returns the time data from when the timer is cleared until output
is started. This command is effective for ASCII data output format. Refer to
“FMT”
on page 4-59.
Execution
Conditions
The CN command has been executed for the specified channel.
If the output voltage is greater than ±42 V, the interlock circuit must be shorted.
Syntax TDV chnum,vrange,voltage[,Icomp[,comp_polarity[,irange]]
Parameters chnum : Source channel number. The value must be slot number where the
module has been installed. Integer expression. 1 to 8 (1 and 5 are not
available for HPSMU). See
Table 4-1 on page 4-9.
vrange: Ranging type for voltage output. Integer expression. The output range
will be set to the minimum range that covers voltage value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
Table 4-4 on page 4-13.
voltage: Output voltage value (in V). Numeric expression. See Table 4-6 on
page 4-14.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Icomp: Current compliance value (in A). Numeric expression. See Table 4-6 on
page 4-14. If you do not set Icomp, the previous value is used. 0 A is
not allowed for Icomp.
comp_
polarity:
Polarity of current compliance. Integer expression.
0: Auto mode (default setting). The compliance polarity is
automatically set to the same polarity as voltage, regardless of the
specified Icomp. If
voltage=0 V, the polarity is set to positive.
1: Manual mode. Uses the polarity of Icomp you specified.
irange: Current compliance ranging type. Integer expression. The compliance
range will be set to the minimum range that covers Icomp value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
Table 4-5 on page 4-13.
Example
Statements
OUTPUT @E5270;"TDV 1,0,20,1E-6,0,15"
ENTER @E5270 USING
"#,5X,13D,X";Time

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals