Keysight PXIe Chassis Family User Guide 141
Updating Chassis Firmware Chassis Maintenance
Viewing the Chassis Revision Information Using the IVI Drivers
The chassis revision string can be viewed using the IVI.NET and IVI-C drivers as
follows:
IVI.NET: Use the Instrument Firmware Revision property.
IVI-C: Use the KTMPXICHASSSIS_ATTR_INSTRUMENT_FIRMWARE_REVISION
attribute.
Updating Chassis Firmware
To determine if there is a later revision of chassis firmware available, perform the
following steps:
1 Go to any of the chassis websites and click the Visit Technical Support tab.
www.keysight.com/find/M9010A
www.keysight.com/find/M9018B
www.keysight.com/find/M9019A
2 Under Technical Support, click the Drivers, Firmware & Software tab. If there are
chassis firmware updates available, they will be listed under this tab, and can
be viewed by clicking Firmware Update under Refine the List.
3 Compare your chassis firmware revision number to the list of chassis firmware
revision numbers that are available. If there are later revisions available,
Keysight recommends installing the latest version.
To install the latest chassis firmware, perform these steps:
1 Click the link to the latest (or desired) firmware update and follow the
instructions provided to download and install the firmware on your chassis.
2 Power cycle the chassis after the firmware installation is complete, and verify
that your host controller PC can communicate to the chassis.