EasyManua.ls Logo

Keysight U1252B - Table 4-1 Available Setting Options in Setup Mode

Keysight U1252B
722 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Changing the Default Setting 4
Keysight U1251B and U1252B User’s and Service Guide 103
Table 4-1 Available setting options in Setup mode
Menu item Available setting options
Default factory setting
Display Description Display Description
rHoLd
[a]
Refresh Hold
OFF
Enables Data Hold
(manual trigger)
500
100–1000
Sets the variation count
that determines Refresh
Hold (auto trigger)
FiLtE DC filter On, OFF
Enables DC filter when
set to On
OFF
bAtt Battery voltage 7.2 V, 8.4 V
Selects battery voltage
to either 7.2 V or 8.4 V
7.2 V
rESEt Reset dEFAU
Enables the reset of
factory settings by
pressing and holding
for more than
1 second
dEFAU
Print Print On, OFF
Enables auto send of
data to the PC
continuously when set to
On
OFF
ECHO Echo On, OFF
Enables the return of
characters to the PC
when set to On
OFF
dAtAb Data bits 7-bit, 8-bit
Sets the data bit length
for the remote
communication (remote
control with PC)
8-bit
PArtY Parity check En, Odd, nOnE
Sets the even, odd or no
parity check for the
remote communication
(remote control with PC)
nOnE
bAUd Baud rate
2400 Hz, 4800 Hz, 9600
Hz, 19200 Hz
Sets the baud rate for
the remote
communication (remote
control with PC)
9600 Hz

Table of Contents