Making Measurements 2
Keysight U1251B and U1252B User’s and Service Guide 63
Audible continuity
In the range of 0–500 , the beeper will sound if the resistance value falls below
10 . For other ranges, the beeper will sound if the resistance falls below the
typical values indicated in Table 2-3 below.
Conductance
Set up the multimeter to measure conductance as shown in Figure 2-9. Probe the
test points and read the display.
The conductance measurement enables the measurement of very high resistance
of up to 100 G.
As the high-resistance readings are susceptible to noise, you can capture the
average readings by using the Dynamic Recording mode. Refer to the section
“Dynamic Recording” on page 78 for more information.
Table 2-3 Audible continuity measurement range
Measurement range Beeper sound threshold
500.00 < 10
5.0000 k < 100
50.000 k < 1 k
500.00 k < 10 k
5.0000 M < 100 k
50.000 M < 1 M
500.00 M < 10 M