3 Functions and Features
82 Keysight U1251B and U1252B User’s and Service Guide
Figure 3-3 Refresh hold mode operation
The HOLD sign
flashes until the
measured value is
stable and updated
Press
Measurement is
updated
to return
Press
Variation counts setting
NOTE
For voltage and current measurements, the holding value will not be updated
if the reading is below 500 counts.
For resistance and diode measurements, the holding value will not be
updated if the reading is in “OL” (open state).
The holding value may not be updated when the reading does not reach a
stable state for all measurements.