Performance Tests and Calibration 6
Keysight U1251B and U1252B User’s and Service Guide 171
3
Press button to go to
the V mode
5 V 0.3 V, 1 kHz N/A 0.3000 V
3 V, 1 kHz N/A 3.0000 V
3 V, 20 kHz * N/A 3.0000 V
50 V 3 V, 1 kHz N/A 03.000 V
30 V, 1 kHz N/A 30.000 V
30 V, 20 kHz * N/A 30.000 V
500 V 30 V, 1 kHz N/A 030.00 V
300 V, 1 kHz N/A 300.00 V
300 V, 20 kHz * N/A 300.00 V
1000 V 30 V, 1 kHz N/A 0030.0 V
300 V, 1 kHz N/A 0300.0 V
300 V, 20 kHz * N/A 0300.0 V
4 Turn the rotary switch to the
position
Short Dual banana plug
with copper wire
short between 2
terminals
SHort SHort
50 mV 30 mV 30.000 mV 30.000 mV
500 mV 300 mV 300.00 mV 300.00 mV
1000 mV 1000 mV 1000.0 mV 1000.0 mV
5
Press button to go to
the mV mode
50 mV 3 mV, 1 kHz 03.000 mV 03.000 mV
30 mV, 1 kHz 30.000 mV 30.000 mV
30 mV, 20 kHz * 30.000 mV 30.000 mV
500 mV 30 mV, 1 kHz 030.00 mV 030.00 mV
300 mV, 1 kHz 300.00 mV 300.00 mV
Table 6-4 Adjustment table (continued)
Step Test Function Cal Range Input
Cal Item
U1251B U1252B